Abstract
The width of metastable zone of unseeded potassium nitrate, potassium sulphate and ammonium sulphate solutions, saturated at 30 °C, was determined by means of an ultrasonic measuring technique at three different cooling rates. The change of ultrasonic velocity and temperature in solutions during a cooling and heating cycle was controlled by using a protected sensor. The results were compared with those of the unprotected sensor taken from the literature for the chosen substances. The change of ultrasonic velocity of supersaturated ammonium sulphate solutions in the presence of crystals was measured at constant temperature by means of a protected sensor. It was found that the protected sensor was reliable and sensitive enough to determine the change of ultrasonic velocity and temperature in both solution and slurry.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Chemical Engineering & Processing: Process Intensification
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.