Abstract

Calibration of temperature modulated calorimeters (TMC) is essential because it is often not possible to calculate heat capacity and sample temperature directly from the quantities measured. The question arises whether temperature calibration used in conventional calorimetry is also valid for TMC’s. To prove this, a well defined transition is needed which does not disturb the temperature profile within the sample during modulation. The response of the whole system (calorimeter, sample, transition) must be linear. Some low energy liquid crystal transitions behave in this way, under certain conditions, and can be used for temperature calibration in TMC’s. The procedure is demonstrated for the nematic to smectic-A transition in a cyanobiphenyl liquid crystal (8OCB). The applicability of 8OCB for temperature calibration is shown for temperature modulated DSC, AC calorimeter, 3 ω method and photo-acoustic method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.