Abstract

Main textThe APMP/TCRI Dosimetry Working Group performed the APMP.RI(I)-K8 key comparison of the reference air kerma rate for high-dose-rate (HDR) 192Ir brachytherapy sources between 2016 and 2018. Eight laboratories, including seven national metrology institutes (NMIs), took part in the comparison. Two commercial well-type chambers were used as transfer instruments and circulated among the participants. The results showed that the maximum difference between the participants and the BIPM, evaluated using the comparison data of the linking laboratories NMIJ and NRC, was less than 4%. The degrees of equivalence for the participants were calculated, so that results for the six non-linking laboratories may be included in the Key Comparison Database (KCDB).To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/.The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call