Abstract
The aim of this study was to determine the amount of apically extruded debris during root canal preparation with ProTaper Gold (PTG) (Dentsply Maillefer, Ballaigues, Switzerland), ProTaper Universal (PTU) (Dentsply Maillefer), ProTaper Next (PTN) (Dentsply Maillefer), and RECIPROC (VDW, Munich, Germany) in curved canals. Eighty human mandibular molar teeth were divided into 4 groups according to the system used for root canal preparation: PTG, PTU, PTN, and RECİPROC groups. The mean weight of extruded debris was statistically analyzed using one-way analysis of variance and post-hoc Tukey's test (α = 0.05). The PTU was associated with significantly more extruded debris than the PTG, PTN, and RECIPROC (p<0.05). The PTG was associated with significantly more extruded debris than the PTN and RECIPROC (p<0.05), whereas no significant difference was found between the RECIPROC and PTN (p>0.05). All systems caused apical debris extrusion during curved canal preparations. However, the PTG, PTN, and RECIPROC systems resulted in significantly less debris extrusion than the PTU system.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.