Abstract

The purpose of this study is to investigate the relationships among cross-sectional diameters, weight and computed tomography (CT) dose descriptors (CTDI and DLP) to identify which is best used as a measure for the establishment of DRLs in CT. Data (gender, weight, cross-sectional diameters, dose descriptors) from 56 adult patients attending for either a CT examination of the abdomen or chest was obtained from two spiral CT units using automatic milliampere modulation. The AP diameter was demonstrated as the main contributing factor influencing the dose in CT (CTDI: r(2) = 0.269, p-value < or =0.001; DLP: r(2) = 0.260, p-value < or =0.001) since it has a greater correlation with radiation dose than body weight and can thus be its substitute in dose-reduction strategies and establishment of DRLs. The advantages of using the AP diameter are that it can easily be measured prior to scanning or retrospectively from previous CT images. However, further studies on the practicality of this approach are recommended.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.