Abstract

To report a new technique for diameter correction in patients after thoracic endovascular aortic repair (TEVAR) with large stent-grafts using the Vascutek Siena 4-branch collared prosthesis (Vascutek Terumo, Inchinnan, Scotland, UK) in patients undergoing thoracoabdominal (TA) aortic replacement. Within a 24-month period, 39 patients underwent TA replacement at our center. Of these, 6 patients had undergone previous TEVAR with large stent-grafts (42 to 48 mm diameter). Indications for TEVAR were aneurysm formation in 3 patients and residual type B aortic dissection (status post repair of type A dissection) in 3 patients. In these patients, the Vascutek Siena 4-branch prosthesis was used in a reversed fashion, removing the elephant trunk portion and using the sewing collar for diameter correction at the anastomotic site. No 30-day mortality and no stroke or symptomatic spinal cord ischemia was observed. Median diameter of the TEVAR graft at the anastomotic site was 44 mm. Median size of the Siena graft used was 26 mm. Diameter correction was successfully achieved in all patients by tailoring the sewing collar of the Siena 4-branch prosthesis to the individual need according to the diameter of the distal end of the stent-graft. Using the sewing collar of the Vascutek Siena 4-branch prosthesis for diameter correction in patients undergoing TA replacement after previous TEVAR with large stent-grafts adds a useful adjunct to the armamentarium of options in a growing patient population.