High-resolution scanning transmission electron microscope (STEM) images in the annular dark field (ADF) imaging mode approaching the theoretical point-to-point resolution limit are presented. The ADF images were obtained from a high Tc superconducting YBa2Cu3O7−x thin-film specimen at 100 kV. The 1.9 Å resolution lattice image, which is the smallest lattice spacing in the specimen, corresponds to the minimum resolvable spatial frequency with 5% contrast in the contrast transfer function for annular dark field, and is smaller than the resolution limit given by the Rayleigh criterion. This demonstrates that STEM ADF imaging can have a resolution approximately 40% better than that of the bright field conventional transmission electron microscope (CTEM) imaging at Scherzer condition.