Abstract

The aim of the present work was to test the combination of non-esterified fatty acid (NEFA) isolation using fumed silicon dioxide with capillary gas-chromatography (C-GC) with splitless injection for the analysis of NEFAs in human plasma. Injection volume, solvent re-condensation and split purge flow-rate were the parameters evaluated for the analysis of fatty acid methyl esters by C-GC. The use of a solvent re-condensation technique, associated with 1.0 µL injection and a split purge flow rate of 80 mL/min resulted in satisfactory analysis of NEFAs. Fourteen fatty acids were identified in plasma samples, ranging from 2.03 to 184.0 µmol/L. The combination of both techniques proved useful for routine analyses of plasma NEFAs.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.