Abstract

In preclinical studies, high-throughput positron emission tomography (PET) imaging, known as simultaneous multiple animal scanning, can reduce the time spent on animal experiments, the cost of PET tracers, and the risk of synthesis of PET tracers. It is well known that the image quality acquired by high-throughput imaging depends on the PET system. Herein, we investigated the influence of large field of view (FOV) PET scanner on high-throughput imaging. We investigated the influence of scanning four objects using a small animal PET scanner with a large FOV. We compared the image quality acquired by four objects scanned with the one acquired by one object scanned using phantoms and animals. We assessed the image quality with uniformity, recovery coefficient (RC), and spillover ratio (SOR), which are indicators of image noise, spatial resolution, and quantitative precision, respectively. For the phantom study, we used the NEMA NU 4-2008 image quality phantom and evaluated uniformity, RC, and SOR, and for the animal study, we used Wistar rats and evaluated the spillover in the heart and kidney. In the phantom study, four phantoms had little effect on imaging quality, especially SOR compared with that for one phantom. In the animal study as well, four rats had little effect on spillover from the heart muscle and kidney cortex compared with that for one rat. This study demonstrated that an animal PET scanner with a large FOV was suitable for high-throughput imaging. Thus, the large FOV PET scanner can support drug discovery and bridging research through rapid pharmacological and pathological evaluation.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call