Abstract

We performed analytical validation of an automated ferric-reducing ability of plasma (FRAP) assay in the serum of dogs. Intra- and interassay precision, accuracy, detection limit, and effects of hemolysis and lipemia were evaluated. Intra- and interassay coefficients of variation were <1% and <13%, respectively. The assay showed a high correlation with a FRAP assay described previously, and results were linear when serial sample dilutions were tested. The detection limit was lower than the values observed in sera from healthy dogs; decreased serum FRAP was found in dogs with leishmaniosis. Lipemia and hemolysis caused a significant increase in the results of the assay.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.