Abstract

The use of arrays of small discrete detectors and thin slices to achieve a high isotropic spatial resolution in positron emission tomography (PET) results in systems with a low ring sensitivity. In a multi-ring system, the overall sensitivity can be considerably improved by removing the interslice collimators to make full use of all cross slice coincidences, but this is achieved at the expense of increased scatter and accidental rates in the image. For imaging of small laboratory animals (diameter of 10 cm or less), the relieved burden of scatters, and to some extent accidentals, suggests that volumetric imaging may be of particular value. In order to evaluate the performance to be expected from a small animal PET scanner (10 cm diameter field) with and without the interplane collimators, the incident event rates for singles (unscattered and single scattered) and true, scatter and accidental coincidences were evaluated analytically. The performance was evaluated for various source sizes and activities. It is concluded that the volumetric configuration is more advantageous than the multislice configuration to image small animals.