Abstract

The analytical performance of the Tandem-R free PSA assay available from Hybritech Inc. was evaluated. Comparison of recoveries of purified free (unbound) prostate-specific antigen (PSA) diluted in female serum in the Tandem-R free PSA assay and the Tandem-R (total) PSA assay demonstrated a link in calibration between the assays and an accurate determination of percent free PSA. The cross-reactivity of the assay to purified PSA-alpha 1-antichymotrypsin was determined to be < 1%. The minimum-detectable concentration was < 0.05 microgram/L. The within-run and between-day CVs were < or = 5% for samples with > 0.3 microgram/L free PSA. Dilution and recovery showed no significant deviations from linearity across the assay range. The assay was insensitive to interference from blood components. The Tandem-R free PSA kit was shown to be an accurate, precise, and reliable assay for the measurement of free PSA.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.