A reverse-phase high-performance liquid chromatography (RP-HPLC) method for the simultaneous estimation of Pyridoxamine dihydrochloride and Acetylcysteine in the marketed formulation is developed. Chromatography carried out at 30oc temperature on Agilent Zorbax Bonus-RP (250 x 4.6 mm, 5 µ) coloum. Coloum using a mobile phase 0.1% trifluroacetic acid in water: acetonitrile (80:20v/v) with flow rate 1ml/min (DAD scan at 210nm). Validation parameters such as system suitability, linearity, precision, accuracy are considered as reported International Conference on Harmonization guidelines. The retention times for Pyridoxamine dihydrochloride and Acetylcysteine are 2 min and 3.4 min. The linearity range for Pyridoxamine dihydrochloride and Acetylcysteine is 30-70 µg/ml and 180-420 µg/ml. The %RSD for accuracy was found to be less than 2%. Hence the proposed method was found to be accurate, precise, reproducible, and specific and can be used for simultaneous analysis of these drugs in tablet formulation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of University of Shanghai for Science and Technology
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.