Abstract

The most common method to determine enhanced dynamic wedge factors begins with the use of segmented treatment tables. These segmental dose delivery values set as a function of upper jaw position are the backbone of a calculation process coined the “MU Fraction Approximation.” Analytical and theoretical attempts have been made to extend and alter the mathematics for this approximation for greater accuracy. A set of linear equations in the form of a matrix are introduced here which correct one published extension of the MU Fraction Approximation as it applies to both symmetric and asymmetric photon fields. The matrix results are compared to data collected from a commissioned Varian Eclipse Treatment Planning System and previously published research for Varian linear accelerators. A total enhanced dynamic wedge factor with excellent accuracy was achieved in comparison to the most accurate previous research found. The deviation seen here is only 0.4% and 1.0% for symmetric and asymmetric fields respectively, for both 6MV and 18MV photon beams.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.