Abstract

The analysis of tetracycline, oxytetracyline, chlortetracycline and doxycycline by high-performance liquid chromatography with pulsed amperometric detection using an anodized boron-doped diamond thin film (BDD) electrode is originally reported. The analyses were carried out using the mobile phase, phosphate buffer (0.01 M, pH 2.5)-acetonitrile (80:20; v/v), on a C18 column (250 x 4.6 mm i.d., 5 microm) at a flow rate of 1.0 mL/min. The optimal PAD waveform parameters at the anodized BDD were 1.5 V (versus Ag/AgCl) detection potential (E(det)) for 290 ms (200 ms delay time and 90 ms integration time), 2.0 V (versus Ag/AgCl) oxidation potential (E(oxd)) for 200 ms oxidation time (t(oxd)) and 0.4 V (versus Ag/AgCI) reduction potential (E(red)) for 200 ms reduction time (t(red)). The proposed method showed the simultaneous determination of tetracycline, oxytetracyline, chlortetracycline and doxycycline with a linear range of 0.1 - 100 microg/mL, detection limits of 0.05 - 0.1 microg/mL and recoveries of 70.8 - 96.0%. The application of this method to real samples was demonstrated and validated using a shrimp sample.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.