Abstract

Slurry sampling in combination with ETV-ICP-AES was employed for the direct determination of trace amounts of impurities in silicon dioxide and silicon nitride. The ETV device consisted of a double layer tungsten coil in a quartz apparatus. Spectral interferences and background emission caused by tungsten ablation of the coil were reduced by coating the coil with tungsten carbide. The background was measured either with a high-purity sample, the suspension medium or close to the analyte emission line, depending on matrix and analyte, or it was calculated using relative emission intensities of tungsten. The concentrations of Al, B, Be, Ca, Cd, Co, Cr, Cu, Fe, Mg, Mn, Ni, Pb and Zn were measured simultaneously, whereas K and Na were determined in the sequential mode. Calibration was performed using the standard additions method. The accuracy was checked by comparison with the results of independent methods. Limits of detection between 0.035 (Mg) and 130 µg g–1 (B) and between 0.01 (Be, Mg) and 34 µg g–1 (B) were achieved in silicon dioxide and silicon nitride, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.