Abstract

An improved method of saxitoxin analysis in urine using continuous-flow fast-atom bombardment mass spectrometry was developed. Parameters studied were matrix composition, matrix flow, temperature of probe tip, probe-tip design and sample extraction. Optimal detection was obtained using the following matrix composition: 5% glycerol, 0.5% acetic acid, 0.025% sodium dodecylsulfate, 0.1% polyethylene glycol (PEG) 400 and 0.5% PEG 300; probe-tip temperature: (approximately 55 degrees C); flow rate: 5 or 8 microL per min.; probe tip: Olson-Hogge design. The STX standard was detected at 200 pg with signal-to-noise ratio of 11. The percent recovery of saxitoxin from human urine after clean-up on a weak cation exchange column was 75%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.