Abstract
Abstract : An improved method of saxitoxin analysis in urine using continuous Flow Fast Atom Bombardment Mass Spectrometry (CF/FAB/MS) was developed. Parameters studied were matrix composition, matrix flow rate, temperature of probe tip, probe tip design and sample extraction. Optimal detection was obtained using the following matrix composition: 5% glycerol, 0.5% acetic acid, 0.025% sodium dodecylsulfate, 0.1% polyethylene glycol (PEG) 400 and 0.5% PEG 300; probe tip temperature: (-55 C); flow rate: 5 or 8 microliter per min; probe tip: Olson-Hogge design. The STX standard was detected at 200 picograms with signal-to-noise ratio of 11. The percent recovery of saxitoxin from human urine after cleanup on a weak cation exchange (CBA Bond-Elut Analytichem Inter. Varian, Anaheim, CA) column was 75%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.