Abstract

This study compared the removal torque between injection molded zirconia implants and titanium implants with resorbable blast media (RBM) surfaces in beagle humeri. Fifteen screw-shaped implants were classified into 3 groups; titanium implant with RBM surface (Group RT), injection molded zirconia implant (Group Zr) and injection molded zirconia implant with sand-blasted surface (Group ZrS). Implants were inserted into beagle humeri. After 12 weeks, removal torque values were measured. The Zr group has a slightly higher removal torque value than the RT and ZrS groups but there were no significant differences among groups. Zirconia implants shows a similar removal torque to RBM titanium implants. This in vivo study showed injection molded zirconia implants could be an alternative to RBM titanium implants in terms of removal torque.