Abstract
This paper describes the scattering corrections needed when neutron detectors are calibrated with a neutron point source at the center of a calibration room. The independence of scattering value from the geometric shape of the room was studied, which for more confidence in this case, two sets of rooms with same inner surface area were evaluated. The parameters that relate the air scattering part (A) and room-return part (R) with the additional contribution from scattered neutrons have been calculated for neutrons whose energy goes from 10-8 to 20MeV. These parameters were calculated using Monte Carlo method for 150, 200, 300, 407, 500, 800 and 1000cm-radius spherical cavity containing air. In the calculations, monoenergetic neutron sources were placed at the center of cavity, and then neutron fluences were determined at several distances of source to detector along the spherical cavity radius. The parameter R has been fitted as a function of surface area (or radius) of spherical room, and the related coefficient has been calculated for 16 monoenergetic neutron sources. It may produce a reasonable estimate (with difference <10%) for the contribution of walls scattering in any geometry of a calibration room.