Abstract
To investigate keratometric measurements according to axial length in an aged population. Patients requiring cataract surgery with keratometric measurements from four different ophthalmic devices (autorefractor/keratometer, Scheimpflug imaging, corneal topography/ray-tracing aberrometry, and partial coherence interferometry) between January 2016 and March 2021 were reviewed retrospectively. Cases for which four ophthalmic devices were deployed in the same order a day were included in this investigation. The corneal curvature of the flattest and steepest meridian, mean corneal curvature, corneal astigmatism, steepest axis location, and axial length were evaluated. In total, 250 eyes (137 patients) were included in the analysis. A negative correlation was found between mean corneal curvature and axial length, with correlation coefficients of 0.587, 0.592, 0.588, 0.591, 0.588, and 0.562 for autorefractor/keratometer, Scheimpflug imaging, corneal topography/ray-tracing aberrometry, partial coherence interferometry, total corneal refractive power of Scheimpflug imaging, and simulated keratometry of corneal topography/ray-tracing aberrometry measurements, respectively. No statistically significant differences were found for corneal astigmatism according to axial length. In axial length group of less than 26.0 mm, negative correlation was found between axial length and mean frontal corneal curvature while no correlation was found between axial length and corneal astigmatism. All four ophthalmic devices showed good inter-device reliability for mean corneal curvature but not corneal astigmatism.
Highlights
ObjectivesAs this report aimed to examine correlations between axial length and corneal curvature, larger samples with both short eye and long eye would be preferable in future studies to confirm the correlations between these variables
The x axes of symmetry on each of six graphs were 27.73, 26.63, 26.08, 27.88, 25.25, and 25.65 We examined the relationship between axial length and corneal astigmatism
As refractive components change with age, the current study aimed to evaluate correlations between axial length and keratometric biometry in an aged population
Summary
As this report aimed to examine correlations between axial length and corneal curvature, larger samples with both short eye and long eye would be preferable in future studies to confirm the correlations between these variables. We aimed to confirm the inter-device keratometric agreement of four ophthalmic devices and determine the relevance of corneal curvature according to axial length
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.