Abstract

Background and purpose To analyze the interfractional set-up errors and intrafractional organ motions and to define appropriate planning target volume (PTV)- and planning organs at risk volume (PRV)-margins in intensity-modulated radiotherapy (IMRT) for head and neck tumors. Patients and methods Twenty-two patients with head and neck or brain tumors who were treated with IMRT were enrolled. The set-up errors were defined as the displacements of the coordinates of bony landmarks on the beam films from those on the simulation films. The organ motions were determined as the displacements of the coordinates of the landmarks on the images recorded every 3 min for 15 min on the X-ray simulator from those on the initial image. Results The standard deviations (SDs) of the systematic set-up errors ( Σ-INTER) and organ motions ( Σ-intra) distributed with a range of 0.7–1.3 and 0.2–0.8 mm, respectively. The average of the SDs of the random set-up errors ( σ-INTER) and organ motions ( σ-intra) ranged from 0.7 to 1.6 mm and from 0.3 to 0.6 mm, respectively. Appropriate PTV-margins and PRV-margins for all the landmarks ranged from 2.0 to 3.6 mm and from 1.8 to 2.4 mm, respectively. Conclusions We have adopted a PTV-margin of 5 mm and a PRV-margin of 3 mm for head and neck IMRT at our department.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.