Abstract

The response of three types of 4 πγ re-entrant well pressure ionization chambers filled with argon or nitrogen were calculated by means of the EGS4 Monte Carlo code as a function of photon energy and compared with experimental results. Excellent agreement was found between calculations and experimental results. This code was also used for the evaluation of corrections required in the change of source conditions such as the displacement of source position. In the medium and high energy region, the chamber response increased when the source was moved to off-center position from center whereas the response decreased in low energy region. This phenomena can be explained by a change ofgeometry and the absorption of photons in an inner cylindrical absorber tube of the chamber.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.