Abstract

End-tidal gas monitoring has become standard of care during inhaled general anaesthesia. We studied the performance of a new side stream gas monitor the ISA multi-gas monitor. The performance was studied at constant low flow of calibration gas and end-tidal anaesthetic measure was studied during routine day case anaesthesia. Pair wise readings of end-tidal halogenated anaesthetic concentration were recorded during low flow anaesthesia. Performance was found to be high; all calibration gas measures were within 0.1 vol% deviation. During routine anaesthesia mean bias was -0.036 vol% and 93 out of 97 pair-wise readings were within the agreement limits as compared to the reference Datex instrument.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.