Abstract

Our patient had recurrent syncope due to ventricular tachycardia (VT) after one year of VVI Pacemaker implantation. He had pacemaker pocket infection for which new pacemaker was implanted on opposite side but old lead was not explanted completely. Flouroscopy showed redundant loop of old lead in right ventricular inflow which was snared out subsequently. He never had syncope or VT after that.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.