Abstract

The layout of the optics for a new small angle x-ray scattering instrument at the DORIS-III wiggler beamline BW4 was investigated by means of ray tracing and finite element calculations. The absolute intensity distributions at different locations of the beamline, especially in the focal plane, have been computed. The first crystal of a double crystal monochromator is the optical element that is most strongly influenced by thermal load. Thus, the distortion of this crystal surface due to thermal load was taken into account as well as the intended cooling system. The results show that the photon flux density at the focus is decreased by a factor of 0.2 due to this distortion.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.