Abstract
An on-line determination of chlorine dioxide (ClO2) in potable water using chlorophenol red (CPR) by gas diffusion flow-injection analysis (FIA) was investigated in the presence of various chlorinated species that can occur under normal water-treatment conditions. A gas diffusion membrane is used to separate the donor (sample) stream from the acceptor (detecting) stream (the donor stream transports the sample stream to the membrane separate device, and the acceptor stream collects all of the penetrated analytes and transports quantitatively to the detector) and makes it possible for this method to eliminate interference from metal ions, as well as other oxychlorinated compounds such as chlorite and chlorate. The system is more selective for chlorine dioxide than chlorine. The linear range of ClO2 concentration is 0–0.5 mg·mL−1 with a detection limit of 0.02 μg·mL−1 (S/N = 3) and a sampling frequency of 50 h−1. © 1999 John Wiley & Sons, Inc. Lab Robotics and Automation 11: 157–161, 1999
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.