Abstract
To evaluate patient radiation exposure for Diagnostic Coronary Angiography (DCA) and Percutaneous Cardiac Intervention (PCI) performed by different operators. Retrospective (n=160) and prospective (n=62) data for DCA (n=179) and PCI (n=43) examinations performed by interventional cardiologists (n=3) using the same imaging equipment were reviewed. The operator with consistently low diagnostic reference levels (DRLs) was interviewed for their personal perceptions upon operator training. Retrospective Median [IQR] DAP was 18.8 [11.8-31.6] and 50.7 [35.3-85.6] Gy.cm2 for DCA and PCI, respectively. Prospective Median [IQR] DAP for DCA and PCI was 7.9 [5.2-10.6] and 15.9 [10.0-17.7] Gy.cm2, respectively. DRLs were within Irish and European DRLs; however, significant inter-operator variability (p<.001) was identified. Radiation exposure in Interventional cardiology is highly operator dependent; further research is warranted in standardization of operator training with evolving technologies.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.