Abstract
Abstract Due to the lack of high voltage and low frequency pulse transmission channels, commercial high frequency IVUS imaging system cannot meet the hardware requirements of ultrasound angiography. Therefore, it is of great research significance to develop a dual-frequency IVUS rotary imaging system with high integration, high sampling rate and sampling accuracy. In this work, an integrated dual-frequency IVUS system with interleaved sampling and parallel signal processing based on FPGA was developed. The central frequency of low-frequency channel is up to 10MHz, and the peak voltage range is 80-200V. The central frequency of the high-frequency channel is up to 60MHz, and the voltage peak-to-peak range is 40-80V. The acquisition channel is 500MHz and 12bit, which is realized by interleaved sampling of two ADC. The motor controlled ultrasonic transducer acquires 400-800 lines per turn, and eventually form a 360-degree IVUS hybrid imaging successfully.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.