Abstract

A method of acoustic imaging is discussed that potentially can improve the diagnostic capabilities of medical ultrasound. The method, given the name second order ultrasound field imaging, is achieved by the processing of the received signals from transmitted dual frequency band pulse complexes with at least partly overlapping high frequency (HF) and low frequency (LF) pulses. The transmitted HF pulses are used for image reconstruction whereas the transmitted LF pulses are used to manipulate the elastic properties of the medium observed by the HF imaging pulses. In the present paper, nonlinear propagation effects observed by a HF imaging pulse due to the presence of a LF manipulation pulse is discussed. When using dual frequency band transmit pulse complexes with a large separation in center frequency (e.g., 1:10), these nonlinear propagation effects are manifested as a nonlinear HF propagation delay and a HF pulse distortion different from conventional harmonic distortion. In addition, with different transmit foci for the HF and LF pulses, nonlinear aberration will occur.