In order to extend the operating range of the GEC RF Reference Cell, we developed an inductively coupled plasma source that replaced the standard parallel-plate upper-electrode assembly. Voltage and current probes, Langmuir probes, and an 80 GHz interferometer provided information on plasmas formed in argon, chlorine, and nitrogen at pressures from 0.1 Pa to 3 Pa. For powers deposited in the plasma from 20 W to 300 W, the source produced peak electron densities between 1010/cm3 and 1012/cm3 and electron temperatures near 4 eV. The electron density peaked on axis with typical full-width at half maximum of 7 cm to 9 cm. Discharges in chlorine and nitrogen had bimodal operation that was clearly evident from optical emission intensity. A dim mode occurred at low power and a bright mode at high power. The transition between modes had hysteresis. After many hours of high-power operation, films formed on electrodes and walls of one Cell. These deposits affected the dim-to-bright mode transition, and also apparently caused generation of hot electrons and increased the plasma potential.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.