An indirect drug delivery device driven by piezoelectric pump (IDDDPP) is proposed, which has the advantages of environmental friendliness and low drug waste. The IDDDPP prototype, which consists of a piezoelectric pump and a replaceable drug delivery unit, is manufactured using materials with good bio-compatibility. The prototype with different needles was designed and manufactured. The experimental researches are performed to examine the influence of different needles on the performance of IDDDPP. Experimental results demonstrate that increasing the diameter of the needle increases the maximum output flow rate of IDDDPP. At 160 Vpp and 24 Hz, the maximum flow rate of 0.96 ml min−1 is obtained with the needle of 0.3 mm diameter.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.