Abstract
IntroductionThe study compared Twisted File (TF) with 3 traditionally manufactured systems to determine whether changes in the manufacturing process improved the cyclic fatigue resistance. MethodsFour rotary file systems, (1) ProFile (PF), (2) Mtwo, (3) K3, and (4) TF, were tested in artificial canals with 45° and 90° angles of curvature. Ten instruments each of the 4 file systems were tested in both angles of curvature (n = 10). All instruments had identical size and taper (.06/0.25 tip diameter). ResultsA statistically significant difference (P < .05) was noted between TF and other nickel-titanium instruments in both 45° and 90° angles of curvature. TF showed the greatest mean number of cycles to failure. There was no statistical difference between PF and K3 (P > .05) in both canal curvatures; however, statistically significant difference (P < .05) was observed between Mtwo and the other 2 traditionally manufactured instruments. Mtwo showed the lowest mean number of cycles to failure. ConclusionsUnder the conditions of this study, size .06/0.25 TF was significantly more resistant to fatigue than the other 3 instrument systems produced with the traditional grinding process.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.