Abstract

Ultrasonic pulse echo systems are often limited in range resolution by the bandwidth of the piezoelectric transducer. Significant improvements in the range resolution of such systems can be obtained by minimizing the effects of the transducer's dynamic response on the overall pulse echo process. An approach to minimize the effects of the transducer is developed from linear system and impulse response techniques. In essence, the pulse echo voltage of interest is deconvolved with a pulse echo reference voltage which is obtained from an air/water interface in the near field of the transducer. A computer study of the pulse echo process and the deconvolution process is presented to illustrate the nature of the improvement in range resolution for several cases of interest. Finally, experimental results are presented to illustrate the improvement using commercially available transducers. [Work supported by NIH.]

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call