Abstract

Ultrasonic pulse echo systems are often limited in range resolution by the bandwidth of the piezoelectric transducer. Significant improvements in the range resolution of such systems can be obtained by minimizing the effects of the transducer's dynamic response on the overall pulse echo process. An approach to minimize the effects of the transducer is developed from linear system and impulse response techniques. In essence, the pulse echo voltage of interest is deconvolved with a pulse echo reference voltage which is obtained from an air/water interface in the nearfield of the transducer . A computer study of the pulse echo process and the deconvolution process is presented to illustrate the nature of the improvement in range resolution for several cases of interest. Finally, experimental results are presented to illustrate the improvement using commercially available transducers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.