The design, operation and performance of a selective excitation double Mössbauer (SEDM) spectrometer are described. An innovative drive-mounted conversion electron detector for scattered γ-ray energy analysis provides a three orders of magnitude improvement in counting efficiency over previous SEDM equipment. Simple digital circuitry together with Wissel Mössbauer velocity transducers and electronics supplies energy synchronization for indefinite collection times. Magnetic materials containing resonant Mössbauer nuclei (e.g., Fe57) can now be conveniently studied using SEDM, offering a unique insight into dynamic processes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.