Abstract
Abstract An improved ELISA for the determination of Thyroglobulin(TG) in human serum with differential pulse voltammetry(DPV) has been developed. The enzymatic product 2,2′-diaminoazobenzene(DAA), produced from horseradish peroxidase(HRP) catalyzing the oxidation of H2O2 with orthophenylene diamine(OPD), yielded a sensitive differential pulse voltammetric response at - 0.18V(Vs Ag/AgCL) in pH2.0 0.2mol/L phosphate-0.1mol/L citrate buffer at a gold electrode. The small three-electrode system was directly inserted in the well of an ELISA plate for detection. The detection limit for TG was 3.8 ng/mL, which was about four times lower than that obtained by measuring the absorbance of DAA.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.