Abstract
A distillation apparatus for a continuous-flow method for determination of phenol in water is described. The colorimetric measurement step utilizes the familiar 4-aminoantipyrine method. Phenol in water can be measured at a rate of 10 samples per hour, has a sensitivity of 4.0 × 10 3 μg l −1 per unit absorbance, a limit of detection of 5 μg l −1, and a standard deviation of 1.6 μg l −1 for 20 replicates of a 100-μg l −1 sample.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.