Abstract

Abstract A high-performance liquid chromatographic method using a microbore system consisting of a microbore column (100 mm – 1mm I. D.- 3 μm), microbore pump (60 μl/min) and UV absorbance detector has been developed for the determination of adenosine in rat brain microdialysis samples. UV absorbance of adenosine at 259 nm was used to measure adenosine. The limit of detection (LOD) was 50 pg and the limit of quantitation (LOQ) was 10 ng/ml. The methodology is applicable for the quantitation of adenosine in small volumes (3 – 10 (μl) of microdialysate and does not require sample preparation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.