Abstract

In this paper, an implantable planar inverted-F antenna (PIFA) for an artificial cardiac pacemaker is proposed. The antenna has a simple structure with a low profile and is placed on the top side of the pacemaker. The dimensions of the pacemaker system, including the antenna element, are <TEX>$42{\times}43.6{\times}11$</TEX> mm. When the antenna is embedded in pig tissue, its <TEX>$S_{11}$</TEX> value is -10.94 dB at 403 MHz and the -10 dB impedance bandwidth of the antenna is 6 MHz (399~406 MHz). The proposed PIFA in tissue has a peak gain of -20.19 dBi and a radiation efficiency of 1.12 % at 403 MHz. When the proposed antenna is placed in a flat phantom, its specific absorption ratio (SAR) value is 0.038 W/kg (1 g tissue). Performances of the proposed PIFA is sufficient to operate at the MICS band (402 ~ 405 MHz).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.