Abstract

This paper reports an implantable microsystem capable of recording neural activity simultaneously on 64 channels, wirelessly transmitting spike occurrences to an external interface. The microsystem also allows the user to examine the spike waveforms on any channel with 8 bit resolution. Signals are amplified by 60 dB with a programmable bandwidth from < 100 Hz to 10 kHz. The input-referred noise is 8 ?Vrms. The channel scan rate for spike detection is 62.5 kS/sec using a 2 MHz clock. The system dissipates 14.4 mW at 1.8 V, weighs 275 mg, and measures 1.4 cm 1.55 cm.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.