Abstract
Abstract. Various methods of determining cyclosporin (CyA) levels in patients after kidney transplantation were compared. These included polyclonal antibody (pcAb-), specific and nonspecific monoclonal antibody (S-and NmcAb-) radioimmu-noassays (RIA), and high performance liquid chromatography (HPLC). The results obtained by the various methods when compared showed some correlation but did not correspond. A probable explanation for part of this noncorrespondence is the presence of monodonaily crossreactive metabolites (CyA-M). Another reason was that the concentration of CyA in the standards supplied with the RIA kits was found to be higher than stated.