Abstract

MRI is the preferred staging modality for rectal carcinoma patients. This work assesses the CT-MRI co-registration accuracy of four commercial rigid-body techniques for external beam radiotherapy treatment planning for patients treated in the prone position without fiducial markers. 17 patients with biopsy-proven rectal carcinoma were scanned with CT and MRI in the prone position without the use of fiducial markers. A reference co-registration was performed by consensus of a radiologist and two physicists. This was compared with two automated and two manual techniques on two separate treatment planning systems. Accuracy and reproducibility were analysed using a measure of target registration error (TRE) that was based on the average distance of the mis-registration between vertices of the clinically relevant gross tumour volume as delineated on the CT image. An automated technique achieved the greatest accuracy, with a TRE of 2.3 mm. Both automated techniques demonstrated perfect reproducibility and were significantly faster than their manual counterparts. There was a significant difference in TRE between registrations performed on the two planning systems, but there were no significant differences between the manual and automated techniques. For patients with rectal cancer, MRI acquired in the prone treatment position without fiducial markers can be accurately registered with planning CT. An automated registration technique offered a fast and accurate solution with associated uncertainties within acceptable treatment planning limits.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call