Abstract

This study describes the design, construction and operation of a new type of transmit/receive array using ceramic resonators operating in a transverse electromagnetic (TE) mode. Single element function and performance at 298.1MHz (7T) are analyzed and compared to a lumped element design loop coil with comparable geometry. The results show that ceramic resonators working in the TE01δ mode configuration produce similar efficiency, defined as the transmit magnetic field (B1+) per square root of the specific absorption rate (SAR), to conventional surface coils. An array consisting of eight ceramic elements was then designed to operate in transmit/receive mode. This array was driven via power/phase splitters by two independent transmit channels and functional cardiac images were produced from a number of healthy volunteers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.