Abstract

This paper considers the applicability of effective conductivity measurements for monitoring physiological and/or pathological phenomena induced by ischemia in the myocardium. The sensitivity of a probe, calculated by means of the finite element method, to changes in the conductivity of the tissue examined is defined for this purpose. Probes developed by Schafer and collaborators (1995) and in our own departments have been examined on the basis of this sensitivity. Theoretical results were verified experimentally using a tank, enlarged models of the probes, and a specially developed electronic circuit. It follows from this study that the probe developed by Schafer et al. is characterized by positive and negative sensitivity. This can lead to misinterpretation of the measurements obtained. In contrast, the sensitivity of our probe is dominantly positive. An example of the in vivo result obtained during experimentally induced ischemia in a swine heart is included.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.