Abstract
AbstractAn automatic sampling device has been developed for capillary zone electrophoresis. The sample is introduced to the column electrokinetically by rapidly exchanging buffer for sample in a narrow channel drilled in a plexiglass block. The autosampler is capable, under computer control, of performing multiple sample injections from a large volume sample source (such as a reaction vessel) as the sample concentration changes, and thus presents the possibility of analyzing time‐varying processes by CZE. Peak area reproducibility in electropherograms obtained after use of the sampler is less than 1% and efficiency is more than 2.2 × 105 theoretical plates.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.