Abstract

An automatic method for the determination of carcinogenic 1-nitropyrene in extracts from automobile exhaust particulate matter has been developed. Organic matter was extracted from particulate matter and the extract was concentrated. Then the extract was injected into a two-dimensional (2D) high-performance liquid chromatograph (HPLC) with a fluorescence detection system. To improve the sensitivity a large sample volume (300 microl) was applied to the HPLC system using the double on-column focusing technique. To achieve full automation and high selectivity a reduction column packed with platinum black was introduced after the 2D HPLC system using a C18 column as the primary column and a pentabromobenzyl column as the secondary column. This HPLC system could determine 0.01-3,000 ng ml(-1) of 1-nitropyrene in an extract from diesel and gasoline exhaust particulate matter at 40 min intervals.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.