Abstract
A routine procedure in which the Technicon Auto Analyzer is used for the determination of boron in plant tissue with azomethine H is described. The preparation of the reagent as the condensation product of H-acid and salicylaldehyde is given. The reagent is highly sensitive to boron, and only copper, iron and aluminium interfere. These interference effects are eliminated with an EDTA solution. The procedure is performed in aqueous medium thus simplifying its adaptation to an automated system and its use for routine determination.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.