Abstract
A computer controlled [ 15O]H 2O injection system has been designed for multiple bolus injections during PET cerebral blood flow studies. The system is housed in a lead safe, and radiation exposure to nearby personnel is low; we have measured rates of approx. 100 μSv/h (10 mR/h) during 2 min of [ 15O]H 2O production and < 50 μSv/h (< 5 mR/h) after 3.7 GBq (100 mCi) has been produced. Reproducibility studies show that the injected activity varies with a coefficient of variation of 2–3%.