Abstract

AZ31 magnesium coronary stents were studied through a manufacturing process chain involving laser cutting, acid pickling, and dip coating. The purpose of this study was to evaluate surface thickness and geometrical dimensions of stents after processing. Stents were dip coated in a solution using PCL with 1% of TiO2. Additionally, AZ31 coronary stents were dynamically tested using a degradation system based on peristaltic pumps. Our results indicate that coated stents degraded slower than AZ31 uncoated control stents. After 4 weeks of dynamic degradation under flowing Hank’s solution, coated stents lost only ∼9% in weight while uncoated stents lost ∼27% in weight. Stents were qualitatively evaluated after four weeks of degradation. Our results demonstrate the formation of micro-pores after one and two weeks of degradation for coated stents. Lamination was observed after three weeks of degradation, meanwhile, uncoated stents resulted with notches and an irregular surface caused by degradation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.